Specifications
The following performance specifications conform to ISA-RP-37.2 (1964) and are typical values, referenced at +75˚F (+24˚C), 4 mA and 100 Hz,
unless otherwise noted. Calibration data, traceable to National Institute of Standards and Technology (NIST), is supplied
Dynamic characteristics Units
Charge sensitivity
Typical pC/g 2.8
Minimum pC/g 2.2
Frequency response See typical amplitude response
Resonance frequency kHz 21
Amplitude response [1]
±5% Hz 1 to 4000
±1 dB (ref) Hz 0.1 to 6000
Temperature response See typical curve
at -67°F (-55°C) max/min % -12 / 0
at +350°F (+177°C) max/min Hz 20 / 0
Transverse sensitivity % 5
Amplitude linearity % 1
Per 500g, 0 to 2000 g
Electrical characteristics
Output polarity Acceleration applied in the direction of the axis arrow produces positive output
Resistance 10
Resistance at +350°F (+177°C) 5
Isolation 10
Signal ground to each signal ground and to mounting surface
Capacitance pF 400
Grounding Each sensor is isolated from the anodized aluminum case
Environmental characteristics
Temperature range -67˚F to +350˚F (-55˚C to +177˚C)
Humidity Epoxy sealed, non-hermetic
Sinusoidal vibration limit g pk 1000
Shock limit [2] g pk 2000
Salt spray Will meet ML-E-5272C, para 4.6.1 when used with sealed connector
Electromagnetic sensitivity equiv. g rms/gauss 0.01
Physical characteristics
Dimensions See outline drawing
Weight gm (oz) 15 (0.53)
Case material Aluminum alloy case, hard adonized, nickel alloy sensors
Connector Mates with Endevco 3060 series cable
Mounting torque lbf-in (Nm) 8 (1)
Calibration
Supplied:
Charge sensitivity pC/g
Capacitance pF
Maximum transverse sensitivity
%
Charge frequency response
% 20 to 4000 Hz
dB thru resonance (Z axis only)
APPLIES TO CALIFORNIA FACILITY
Model 2228C
Piezoelectric accelerometer
Continued product improvement necessitates that Endevco reserve the right to modify these specifications without notice. Endevco maintains a program of constant surveillance over
all products to ensure a high level of reliability. This program includes attention to reliability factors during product design, the support of stringent Quality Control requirements,
and compulsory corrective action procedures. These measures, together with conservative specifications have made the name Endevco synonymous with reliability.
©Meggitt (San Juan Capistrano), Inc. All Rights Reserved 30700 Rancho Viejo Road, San Juan Capistrano, CA92675 USA
(800) 982-6732 • (949) 493-8181 fax (949) 661-7231 • www.endevco.com • Email: [email protected]om 011911
Product Description 2228C 2228C-R
3060D-120 [3] Cable assembly, three each [3], 10 ft Included Optional
EH156 4-40 x 5/8 socket head cap screws Included Included
two each
EHW53 No. 4 flat washer, two each Included Included
EHM464 Hex wrench Included Optional
2771C In-line charge convertor Optional Optional
133 Signal conditioner Optional Optional
2775B Signal conditioner Optional Optional
4990A-1 OASIS 2000 computer-controlled system Optional Optional
6634C Signal conditioner Optional Optional
Accessories Notes:
1. Low-end response of the transducer is a function of its associated electronics.
2. Short duration shock pulses, such as those generated by metal-to-metal
impacts, may excite transducer resonance and cause linearity errors.
Read TP290 for more details.
3. Flexible cable, such as the supplied 3060D, should be used to minimize
cable-strain errors.
4. Adhesives such as petro-wax, hot-melt glue, and cyanoacrylate epoxy
(super glue) may be used to mount the accelerometer temporarily to the
test structure. An adhesive mounting kit (P/N 31849) is available as an option
from Endevco. To remove an epoxy-mounted accelerometer, first soften the
epoxy with an appropriate solvent and then twist the unit off with the supplied
removal wrench. Damage to sensors caused by inappropriate removal
procedures are not covered by Endevco’s warranty.
5. Maintain high levels of precision and accuracy using Endevco’s factory
calibration services. Call Endevco’s inside sales force at 800-982-6732 for
recommended intervals, pricing and turn-around time for these services as
well as for quotations on our standard products.
Vue de la page 0
1

Résumé du contenu

Page 1 - Piezoelectric accelerometer

SpecificationsThe following performance specifications conform to ISA-RP-37.2 (1964) and are typical values, referenced at +75˚F (+24˚C), 4 mA and 100

Commentaires sur ces manuels

Pas de commentaire