All specifications apply over 0 °C to + 55 °C unless otherwise noted
and are covered by the product warranty. The analyzer will meet its
specifications when: it’s within the one year calibration cycle, AUTO
ALIGN [ALL] is selected, stored a minimum 2 hours within the oper-
ating temperature range, turned on for at least 5 minutes, Align Now
RF has been run once every 24 hour period. Characteristics describe
product performance that is useful in the application of th product,
but is not covered by the product waranty. Typical performance is
beyond specifications that 80% of the units exhibit 95% confidence
level over 20 to 30°C not including measurement uncertainty and is
not covered by the product warranty.
Frequency specifications
Frequency range
E4401B
50 9 kHz to 1.5 GHz
75 1 MHz to 1.5 GHz
E4402B 9 kHz to 3.0 GHz
dc coupled (Option UKB) 30 Hz
6
to 3 GHz
ac coupled (Option UKB) 100 kHz to 3 GHz
E4404B
dc coupled 9 kHz to 6.7 GHz
dc coupled (Option UKB) 30 Hz
6
to 6.7 GHz
ac coupled 100 kHz to 6.7 GHz
Band
0 9 kHz to 3.0 GHz
(Option UKB) 100 Hz to 3.0 GHz
1 2.85 GHz to 6.7 GHz
E4405B
dc coupled 9 kHz to 13.2 GHz
dc coupled (Option UKB) 30 Hz
6
to 13.2 GHz
ac coupled 100 kHz to 13.2 GHz
Band N
4
0 1– 9 kHz to 3.0 GHz
0 (Option UKB) 30 Hz
6
to 3.0 GHz
1 1– 2.85 GHz to 6.7 GHz
2 2– 6.2 GHz to 13.2 GHz
E4407B
Internal mixing 9 kHz to 26.5 GHz
dc coupled (option UKB) 30 Hz
6
to 26.5 GHz
ac coupled (option UKB) 10 MHz to 26.5 GHz
Band N
4
0 1– 9 kHz to 3.0 GHz
0 (option UKB) 30 Hz
6
to 3.0 GHz
1 1– 2.85 GHz to 6.7 GHz
2 2– 6.2 GHz to 13.2 GHz
3 4– 12.8 GHz to 19.2 GHz
4 4– 18.7 GHz to 26.5 GHz
External mixing (Option AYZ) 18 GHz to 325 GHz
Agilent E4401B, E4402B, E4404B,
E4405B, and E4407B
Data Sheet
ESA-E Series Spectrum Analyzers
These specifications apply to the Agilent
Technologies E4401B, E4402B, E4404B,
E4405B, and E4407B spectrum analyzers.
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - E4405B, and E4407B

All specifications apply over 0 °C to + 55 °C unless otherwise notedand are covered by the product warranty. The analyzer will meet itsspecifications

Page 2 - Amplitude specifications

2Frequency reference(Option 1D5)Aging ±2 × 10–6/year ±1 × 10–7/yearTemperature stability ±5 × 10–6±1 × 10–8 (20 to 30 °C)Settability ±5 × 10–7±1 × 10–

Page 3 - Overall amplitude accuracy

3Maximum safe input levelAverage continuous power(input attenuator ≥15 dB)E4401B +30 dBm (1 W)E4401B (75 Ω Option 1DP) +75 dBmV (0.4 W)(input attenua

Page 4 - General specifications

4RF input VSWR6(at tuned frequency, 10 dB attenuation)E4401B1 MHz to 1.5 GHz 1.35:1 E4402B9 kHz to 100 kHz

Page 5 - Option specifications

5Data storage (nominal)Internal 200 traces or statesExternal (1.44 MB floppy disk) 200 traces or statesWeight6(without options)E4401B 13.2 kg (29.1 l

Page 6 - Option AYZ external mixing

6Spurious outputHarmonic spursE4401B(0 dBm output)9 kHz to 20 MHz <–20 dBc20 MHz to 1.5 GHz <–25 dBcE4402B/04B/05B/07B(–1 dBm output)20 kHz to 3

Page 7

7Notes1. Frequency reference error = (aging rate x period of time since adjustment + settability + temperature stability).2. Not available in RBW &l

Page 8

Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive,while minimizi