Technical SpecificationsAgilent RF Network Analyzers PNA SeriesThis document describes the performance and features of Agilent Technologies PNA Series
10Description Specification Supplemental informationTest port noise floora300 kHz to 25 MHzb10 Hz IF bandwidth ≤ –115 dBm1 kHz IF bandwidth ≤ –95 dBm2
11Test port input (continued)Description Specification Supplemental informationReference level magnitudeRange ±200 dBResolution 0.001 dBReference leve
12Test port input (continued)Dynamic accuracyAccuracy of the test port input power reading is relativeto the reference input power level. Applies to i
13Test port input (continued)Group delayaDescription Specification Supplemental informationAperture (selectable) (frequency span)/(number of points –
General information (continued)Description Supplemental informationDisplay 21.3 cm (8.4 in) diagonal color active matrix LCD; 640 (horizontal) x 480 (
15a. A third-wire ground is required.Description Supplemental InformationExternal AM input BNC, female; voltage input provides low frequency AM modul
16Measurement throughput summaryCycle time vs. IF bandwidthaInstrument state: preset condition, 201 points, CF = 1 GHz,Span = 100 MHz, correction off,
17Data transfer time (ms)aNumber of points51 201 401 1601SCPI over GPIB(program executed on external PC)b32-bit floating point 4 7 13 4164-bit floatin
18Standard Option 015 PNA Series simplified test set block diagram
19Measurement capabilitiesNumber of measurement channelsUp to four independent measurement channels. A measurement channel is coupled to stimulus resp
2Some definitionsAll specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90minutes after the instrument ha
20Data accuracy enhancementMeasurement calibrationMeasurement calibration significantly reduces mea-surement uncertainty due to errors caused by syste
21System capabilitiesFamiliar graphical user interfaceThe PNA Series analyzer employs a graphical userinterface based on Windows®2000. There are two f
22Configurable test set (Option 015)With the configurable test set option, front panel accessloops are provided to the signal path between thesource o
23AutomationMethodsInternal analyzer executionWrite applications that can be executed from withinthe analyzer via COM (component object model) orusing
24Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value youreceive, while minimi
3Corrected system performanceThe specifications in this section apply for measure-ments made with the PNA Series analyzer with thefollowing conditions
4Corrected system performance with type-N connectorsApplies to PNA Series analyzer, 85032F (Type-N, 50 Ω)calibration kit, and N6314A test port cable u
5Corrected system performance with type-N connectorsApplies to PNA Series analyzer, 85092B (Type-N, 50 Ω) Electronic calibration (ECal) module, andN63
6Corrected system performance with type-N connectorsApplies to PNA series analyzer with Option 015,85032F (Type-N, 50 Ω) calibration kit, and N6314A t
7Corrected system performance with 3.5 mm connectorsApplies to PNA Series analyzer, 85033E (3.5 mm, 50 Ω)calibration kit, and N6314A test port cable u
8Corrected system performance with 3.5mm connectorsApplies to PNA Series analyzer, 85093B (3.5mm, 50 Ω)Electronic calibration (ECal) module, and N6314
9Uncorrected system performanceDescription Specification (dB)300 kHz to 1 MHz 1 MHz to 1.3 GHz 1.3 GHz to 3 GHz 3 to 6 GHz 6 to 9 GHzDirectivity 30 33