Hewlett Packard 3577A 5Hz - 200 MHz
Network Analyzer Specifications
SOURCE
Frequency
Range
5 Hz - 200 MHz
Resolution
0.001 Hz
Amplitude
± 5 X 10
-8
/day, 0 to 55°C
Range
+15 to -49 dBm (1.26 Vrms to 793m Vrms: 2dBV to -62 dBV)
into a 50Ω load
Resolution
0.1 dB
Accuracy
± 1 dB at +15 dBm and 100 kHz. Below +15 dBm, add the greater
of ± 0.02 dB/dB or 0.2 dB
Flatness
1.5 dBp-p from 5 Hz to 200 MHz
Impedance
50Ω; > 20 dB return loss at all levels
RF Output Connector
50Ω Type N female
Sweep Types
Linear, alternate, cw and log frequency; log amplitude
Sweep Time
100 ms/span to 200 ms/span for frequency sweep; 1 ms/step to 16
s/step for amplitude sweep
Sweep Modes
Continuous, single, manual
Trigger Modes
Free run, immediate, line, external
RECEIVER
Full Scale Input Level
-13 dBV from 10 kHz to 200 MHz with internal 20 dB attenuators
ON (0dBm at 50Ω)
INPUT CHARACTERISTICS
Frequency Range
5 Hz - 200 MHz
Inputs
3 (A, B, R)
Input Impedance
Selectable 50Ω with > 25 dB return loss, or 1 MΩ
in parallel with
approximately 30 pF
Input Connectors
50Ω Type N female
Resolution Bandwidth
Selectable 1 kHz, 100 Hz, 10 Hz, or 1 Hz
30 kHz - 200 MHz (50Ω)
30 kHz - 20 MHz (1 MΩ)
Sensitivity (Due to noise and
internal crosstalk between
source and receiver inputs)
Resolution
Bandwidth
Internal 20 dB
Attenuator ON
Internal 20 dB
Attenuator OFF
Vue de la page 0
1 2 3 4 5

Résumé du contenu

Page 1

Hewlett Packard 3577A 5Hz - 200 MHz Network Analyzer Specifications SOURCE Frequency Range 5 Hz - 200 MHz Resolution 0.001 Hz Amplitude ± 5 X 10 -8 /d

Page 2

1 Hz 10 Hz 100 Hz 1 kHz -110 dBm -110 dBm -105 dBm -95 dBm -130 dBm -130 dBm -125 dBm -115 dBm Crosstalk > 100 dB isolation between inputs Electr

Page 3

Display: 0.01 deg/div to 200 deg/div (0.00018 rad/div to 3.49 rad/div) Accuracy (at 100 kHz, 25°C, and Full Scale Input) ± 2.0° Error Input Level Rela

Page 4

Logarithmic and Linear. Polar/Smith Chart Graticule ± 500 deg in 0.001 deg increments Noise Averaging Type Exponentially weighted vector averaging on

Page 5

Save/Recall Front-panel setups can be stored in non-volatile memory locations 1 through 5. Last state is saved when power is removed. Operating Condit