Page 1 - E4980AL Precision LCR Meter

E4980A Precision LCR Meter20 Hz to 2 MHzE4980AL Precision LCR Meter20 Hz to 300 kHz/500 kHz/1 MHzData SheetFully compliant to LXI Class C specificatio

Page 2 - E4980A/E4980AL

10Q accuracy (when Q x De < 1)Q accuracy Qe is given as:Equation 10. Qe = ± (Qx2 × De) (1 Qx × De) Qx Measured Q valueDe R

Page 3 - Basic Specifications

11Example of C-D accuracy calculationMeasurement conditionsTest Frequency: 1 kHzMeasured C value: 100 nFTest signal voltage: 1 VrmsMeasurement

Page 4

12Effect by impedance of DUTTable 14. For impedance of DUT below 30 Ω, the following value is added.Test Impedance of DUTfr

Page 5 - • Level monitor accuracy:

13Short offset ZsTable 16. Impedance of DUT > 1.08 ΩTest Measurement time modefrequency [Hz] SHORT MED, LONG20 -

Page 6 - Measurement display ranges

14Effect of cable length Table 21. When the cable is extended, multiply Yo by the following factor.Test Cable lengthfr

Page 7 - Absolute measurement accuracy

15Calibration accuracy Acal Calibration accuracy Acal is given below.For impedance of DUT on the boundary line, apply the smaller value.Table 23. Impe

Page 8

16Measurement accuracyThe impedance measurement calculation example below is the result of absolute measurement accuracy.Figure 1. Impedance measureme

Page 9 - Relative accuracy

17Compensation functionTable 28. The E4980A provides three types of compensation functions: OPEN compensation, SHORT compensation, and LOAD compensati

Page 10

18Comparator function of list sweep: The comparator function enables setting one pair of lower and upper limits for each measurement point. You can s

Page 11 - Basic accuracy

19The following options are available for the E4980A LCR Meter. Frequency optionsNone of below 20 Hz to 2 MHzOption 030 20 Hz to 300 kHzOption 050

Page 12 - Effect of cable extension

2All specifications apply to the conditions of a 0 to 55 °C temperature range, unless otherwise stated, and 30 minutes after the instrument has been t

Page 13 - Open offset Yo

20Power and DC Bias enhancement specificationIncreases test signal voltage and adds the variable DC bias voltage function.This function is available w

Page 14 - Effect of cable length

21Table 32. Test signal currentRange 0 Arms - 100 mArmsResolution 1 µArms (0 Arms - 2 mArms) 2 µArms (2 mArms - 5 mArms) 5 µArms (5 mArms

Page 15 - Calibration accuracy Acal

22DC bias signalTable 35. Test signal voltageRange –40 V to +40 VResolution Setup resolution: 100 µV, effective resolution: 330 µV ±(0 V - 5

Page 16 - Measurement accuracy

23Table 37. Test signal voltage ≤ 0.2 Vrms (measurement time mode = MED, LONG)DC bias Impedance range [Ω]current range < 100 100 300,

Page 17 - Compensation function

24DCR measurement specificationDC resistance (Rdc) measurement function is available when either option 001, 030, 050, 100, or 200 is installed.DC res

Page 18 - DC bias signal

25Effect of cable length (Short offset)Table 46. The following value is added to Rs when the cable is extended.Cable length1 m 2 m 4 m0.25 mΩ 0.5 m

Page 19 - Other options

26Table 48. Power source Voltage 90 VAC - 264 VACFrequency 47 Hz - 63 HzPower consumption Max. 150 VATable 49. Operating environmentTemperature 0

Page 20 - Test signal

27Figure 4. Dimensions (rear view, with handle and bumper, in millimeters, nominal) Figure 5. Dimensions (front view, without handle and bumper, in mi

Page 21 - Table 32. Test signal current

28Figure 6. Dimensions (side view, with handle and bumper, in millimeters, nominal) Figure 7. Dimensions (side view, without handle and bumper, in mil

Page 22 - Table 36. Test signal current

29Description Supplemental InformationEMC European Council Directive 2004/108/EC IEC 61326-1:2005 EN 61326-1:2006 CISPR 11:2003 +A1:2004 EN 55011:

Page 23 - 100 Ω (nominal)

3Measurement functionsMeasurement parameters• Cp-D, Cp-Q, Cp-G, Cp-Rp• Cs-D, Cs-Q, Cs-Rs• Lp-D, Lp-Q, Lp-G, Lp-Rp, Lp-Rdc1 • Ls-D, Ls-Q, Ls-Rs

Page 24 - DCR measurement specification

30Settling timeTable 51. Test frequency setting timeTest frequency setting time Test frequency (Fm)5 ms Fm ≥ 1 kHz12 ms 1 kHz > Fm ≥ 250 Hz22 ms

Page 25 - Temperature coefficient Kt

31Measurement timeDefinitionThis is the time between the trigger and the end of measurement (EOM) output on the handler interface.ConditionsTable 54 s

Page 26 - Table 50. Storage environment

32Table 55. Measurement time when option 030/050/100 is installed [ms] Measurement time mode Test frequency 20 Hz 100 Hz 1 kHz 10 kH

Page 27 - 22.23712.7

33When DC bias is ON, the following time is added:Table 57. Additional time when DC bias is ON [ms]Test frequency20 Hz 100 Hz 1 kHz 10 kHz 100 kHz

Page 28 - 21.921.9

34Display timeExcept for the case of the DISPLAY BLANK page, the time required to update the displayon each page (display time) is as follows. When a

Page 29 - Environment

35DC bias test signal current (1.5 V/2.0 V): Output current: Max. 20 mAOption 001 (Power and DC Bias enhance): DC bias voltage: DC bias voltage appli

Page 30 - Information

36Table 66. Yo_DCI1 valueDC bias current range Measurement time mode SHORT MED, LONG20 µA 0 S 0 S200 µA 0.25 nS 0.05 nS2 mA 2.5 nS 0.5 nS20 mA

Page 31 - Measurement time

Web ResourcesVisit our Web sites for additional product information and literature.E4980A Precision LCR Meter www.agilent.com/find/e4980aE4980AL Preci

Page 32 - (DC bias: OFF)

Table 1. Trigger delay time Range 0 s - 999 sResolution 100 µs (0 s - 100 s) 1 ms (100 s - 999 s)Table 2. Step delay timeRange 0 s - 999 sRes

Page 33 - Test frequency

5Signal levelTable 6. Test signal voltageRange 0 Vrms - 2.0 VrmsResolution 100 µVrms (0 Vrms - 0.2 Vrms) 200 µVrms (0.2 Vrms - 0.5 Vrms) 5

Page 34 - Display time

6Measurement display rangesTable 10 shows the range of measured value that can be displayed on the screen. For the effective measurement ranges, refer

Page 35 - Maximum DC bias current

7Absolute measurement accuracyThe following equations are used to calculate absolute accuracy.Absolute accuracy Aa of |Z|, |Y|, L, C, R, X, G, B (L, C

Page 36 - DC bias settling time

8G accuracy (when Dx ≤ 0.1)Equation 5. Bx + Da (S) Bx = 2πfCx = 1 2πfLx Dx Measured D valueBx Measured B value (S)Da Absolut

Page 37 - Web Resources

9Relative accuracyRelative accuracy includes stability, temperature coefficient, linearity, repeatability, and calibration interpolation error. Relati