Agilent 34980A Multifunction Switch/Measure Unit Data Sheet• 8-slot mainframe with 19 mix-and-match plug-in modules so you cancreate your own custom c
34980A matrix switch modulesThe 34980A matrix modules are fullcross-point matrices that allow you to connect any row to any column.This is a convenien
11Table 3. Matrix selection table—specifications and characteristics*34931A 34932A 34933AChannels/configurations dual 4x8 dual 4x16 dual 4x88x8 8x16 8
34980A general-purpose switchmodules The 34980A general-purpose switchescan be used to route signals or to control other system devices. Theseswitches
34980A RF and microwaveswitch modulesThe 34980A offers a variety of RFand microwave switch modules—RF multiplexers, SPDT switchingfrom DC to 20 GHz, o
14Table 5. RF and microwave selection table—specifications and characteristics*DC to 3 GHz DC to 20 GHz [3]34941A 34942A 34946A 34947AChannels quad 1x
34945A/34945EXT microwaveswitch/attenuator driverThis module allows you to controlswitches attenuators, and otherdevices external to the 34980A. The 3
34950A 64-channel digital I/O with memory and counter This module can be used to simulateor detect digital patterns. It has eight8-bit digital I/O cha
34951A 4-channel isolated D/A converter with wave-form memory This module has four independent,isolated channels that output DC voltage up to ± 16 V o
34952A multifunction modulewith 32-bit DIO, 2-channel D/Aand totalizer The multifunction module offers theflexibility you need for system control.The
34959A breadboard module Use this module to create your owncustom designs inside the 34980Amainframe. You can control your custom circuits with access
If you use automated test equipmentfor design validation or manufactur-ing, you now have a cost-effective alternative to PXI and VXI test-systemplatfo
2034980A system specifications and characteristics(for complete specifications, see the 34980A Users Guide)DMM accuracy ± (% of reading + % of range)I
21Temperature measurement accuracy ± (% of reading + % of range)Temperature Type 1-year accuracy [1]24 hour Extended range[1]1-year accuracy Temp Coef
22Measurement characteristics with optional internal DMMDC voltageMeasurement method Continuously integrating multi-slope III A-D converterA-D linear
23System specificationsScanning inputsAnalog: 34921A, 34922A, 34923A, 34924A, and 34925A multiplexer channelsDigital: 34950A/52A digital in and tota
24Ordering informationDescription Optional Interconnect Commentsterminal block cables / connector kits34980A Multifunction switch/measure mainframe— N
25AccessoriesY1130A Rackmount kit for 34980A, forward or reverse mountCables/connector KitsY1135A 1.5m 50pin Dsub, M/F twisted pair with outer shield
Agilent Technologies’ Test and MeasurementSupport, Services, and AssistanceAgilent Technologies aims to maximize the valueyou receive, while minimizin
Make system connections easily and quickly with simple, reliable connection options:• Built-in Ethernet, USB 2.0, andGPIB connectivity • Standard IVI
4Modules provide flexible system stimulus and controlSystem control—with analog outputs,open-collector digital outputs, clockgeneration, and isolated
Intuitive front panel with self-guiding menusSet up scan listsSee results on bright, multiline displayScan multiple channels, close specified channel
6The 34980A mainframe holds up to eight plug-in modules.Mix and match them to create a custom system to meetyour switching and system control needs. Y
734980A multiplexer switchmodulesThe 34980A multiplexer modules can be used to connect one of manydifferent points to a single point. You can connect
Multiple multiplexers can connect to the built-in analog buses, allowingyou to scan up to 560 2-wire channelsor 640 1-wire channels in a singlemainfra
9Table 2. Multiplexer selection table—specifications and characteristics*34921A 34922A 34923A 34924A 34925AChannels/configurations 40 2-wire 70 2-wir
Commentaires sur ces manuels