Version03.00January 2006EMI Test Receivers ¸ESIBEMI measurements up to 40 GHz conforming to standardsData sheetOutstanding performanceHigh sensitivity
10 EMI Test Receivers ¸ESIB¸ESIB 7 ¸ESIB 26 ¸ESIB 40Input 2 (receiver mode) 20 Hz to 1 GHzDC voltageDC coupling 0 VAC coupling 50 VRF attenuation 0 d
EMI Test Receivers ¸ESIB 11¸ESIB 7 ¸ESIB 26 ¸ESIB 40Displayed noise fl oor (receiver mode)Linear average (AV) display (preamplifi er off/on)20 Hz to
12 EMI Test Receivers ¸ESIB¸ESIB 7 ¸ESIB 26 ¸ESIB 40Spurious response (f > 1 MHz, without input signal, 0 dB RF attenuation)Receiver mode or span
EMI Test Receivers ¸ESIB 13¸ESIB 7 ¸ESIB 26 ¸ESIB 40Trigger functionsTrigger free-run, line, video, RF, externalDelayed sweepTrigger source free-ru
14 EMI Test Receivers ¸ESIB¸ESIB 7 ¸ESIB 26 ¸ESIB 40Probe power supply +15 V DC, –12.6 V DC and ground, max. 150 mAPower supply and coding connector
EMI Test Receivers ¸ESIB 15¸ESIB 7 ¸ESIB 26 ¸ESIB 40Dimensions (W × H × D) 435 mm × 236 mm × 570 mmWeight 25.1 kg 26.4 kg 27.0 kg1) After 30 days
Certified Quality SystemISO 9001DQS REG. NO 1954 QMCertified Environmental SystemISO 14001DQS REG. NO 1954 UMMore information at www.rohde-schwarz.com
2 EMI Test Receivers ¸ESIBThe ¸ESIB family of EMI test receiv-ers combines the high sensitivity, large dynamic range and selectivity of a high-end te
EMI Test Receivers ¸ESIB 3Pulse weighting using quasi-peak, peak, average and CISPR-average detectors; the detectors operate in parallel and can b
4 EMI Test Receivers ¸ESIBUp to 2 MHz, the ¸ESIB family uses fi xed-tuned fi lters; from 2 MHz to 1000 MHz, the preselection fi lters oper-ate as track
EMI Test Receivers ¸ESIB 5alignment of the mixer. The ¸ESIB consequently features suffi cient inherent noise suppression with respect to rele-vant l
6 EMI Test Receivers ¸ESIBSplit-screen displayCritical emissions can be measured with numeric display of frequency and level as with classic receive
EMI Test Receivers ¸ESIB 7Typical SSB phase noise100 Hz 1 kHz 10 kHz 100 kHz 10 MHz1 MHz Carrier offset–70–80–90–100–110–120–130–140–150–155–60¸ESI
8 EMI Test Receivers ¸ESIBSpecifi cationsSpecifi cations apply under the following conditions: 30 minutes warm-up time at ambient temperature, specifi e
EMI Test Receivers ¸ESIB 9¸ESIB 7 ¸ESIB 26 ¸ESIB 40Preselector (receiver mode)Filters Frequency range Bandwidth (–6 dB)1 <150 kHz 230 kHz fi xe
Commentaires sur ces manuels