10 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Level display (receiver mode) Screen 625 x 500 pixels, max. 2 diagrams analog nume
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 11 Display nonlinearity +20 °C to +30 °C, mixer level ≤–10 dBm RBW ≤100 kHz or channe
12 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Trigger functions Trigger analyzer mode free run, video, external, IF level (mixer le
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 13 Inputs and outputs (rear panel) IF 20.4 MHz BNC female Impedance 50 Ω RBW ≤30
14 R&S®ESU EMI Test Receiver Version 02.00, October 2007 General specifications Display 21 cm LC TFT color display (8.4") Resolution
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 15 R&S®FSU-B9 Tracking Generator, R&S®FSU-B12 Attenuator for Tracking Generator
16 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Modulation Modulation format external I/Q, AM, FM full scale AM, FM, V pp 1 V Input
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 17 R&S®ESU-B24 Low-Noise Preamplifier R&S®ESU8 100 kHz to 8 GHz R&S®ESU26
18 R&S®ESU EMI Test Receiver Version 02.00, October 2007 RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW presel
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 19 RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW pre
2 R&S®ESU EMI Test Receiver Version 02.00, October 2007 CONTENTS BASE UNIT ...
20 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Noise indication (receiver mode) Nominal, calculated from DANL data RF attenuation = 0
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 21 R&S®ESU26: 20 MHz ≤ f < 30 MHz, BW = 9 kHz <–15 dBµV 30 MHz ≤ f < 1 G
22 R&S®ESU EMI Test Receiver Version 02.00, October 2007 DC coupling, RF attenuation ≥10 dB, preselection = ON, preamplifier = ON8 +20 °C to +30
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 23 Recommended extras Designation Type Order No. Headphones 0708.9010.00 Printed man
Certified Quality SystemISO 9001DQS REG. NO 1954 QMCertified Environmental SystemISO 14001DQS REG. NO 1954 UMFor product brochure, see PD 5213.6693.12
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 3 Base Unit Frequency R&S®ESU8 DC coupled 20 Hz to 8 GHz AC coupled 1 MHz to 8 GH
4 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Receiver scan Scan scan with max. 10 subranges with different settings Measurement t
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 5 IF and resolution bandwidths 3 dB bandwidths 10 Hz to 10 MHz in 1/2/3/5 sequence 1
6 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Intermodulation 0 dB RF attenuation, preselection/preamplifier = OFF1 ≤3.6 GHz +13 dBm,
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 7 R&S®ESU26 20 MHz ≤ f < 1 GHz <–152 dBm, typ. –156 dBm 1 GHz ≤ f < 2 GHz
8 R&S®ESU EMI Test Receiver Version 02.00, October 2007 Noise indication (receiver mode) Nominal, calculated from DANL data RF attenuation = 0
Version 02.00, October 2007 R&S®ESU EMI Test Receiver 9 R&S®ESU26/40 20 MHz ≤ f < 30 MHz, BW = 9 kHz <–16 dBµV 30 MHz ≤ f < 50