High-Current DC Loads
4700 LXI SERIES
K 120V, 1kW/200A to 36kW/7200A
K 18 Voltage, Current, Power and
Timing Measurements
K Complex load profile playback
K Air cooling
K Touch-panel manual control
APPLICATION
4700 LXI Series Electronic Loads are designed for testing applications that require high-power and high-current
plus measurement of UUT dynamic performance. The Loads provide exceptional reliability combined with the
inherent simplicity and safety of air cooling. The 4700 Loads are controlled through either a manual touch-panel, a
LabVIEW soft-panel or an emPower™ Test Executive enabled automatic test station. Typical power conversion
products to be tested include higher-power DC supplies, telecom rectifiers, fuel cells and batteries.
Turn-on Time
Settle Time
Peak Settle
Overshoot
max
min
Rise
Time
Turn-on voltage & current measurements
EXTENDED INTERNAL MEASUREMENTS
The 4700 Loads eliminate the need for separate external
instruments and related switching to capture timing and other
dynamic measurements during UUT turn-on and turn-off. This
is accomplished by incorporating the task-essential features of
a DMM, DSO and Transient Generator. With this built-in
capability, the 4700 can rapidly generate engineering-
characterization-like test information at faster test speeds with
the additional benefit of cost savings from fewer measurement
instruments, switches and interconnection wiring.
Power supply turn-on voltage & current
waveforms in Auto Mode
Constant Current
Constant Resistance
Constant Voltage
Current
Voltage
AUTO MODE
The 4700 contains a unique Auto Mode that provides
glitchless automatic switching between CR, CC, CV and CP
limits. Previously challenging tests that are now made
possible include turn-on into a CR load, confirming the
complete V-I curve for Lithium-Ion battery chargers, and
preventing device-under-test blowup should a protection
circuit fail.
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - High-Current DC Loads

High-Current DC Loads4700 LXI SERIESK 120V, 1kW/200A to 36kW/7200A K 18 Voltage, Current, Power andTiming MeasurementsK Complex load profile playback

Page 2 - POWERTOUCH MANUAL CONTROL

Just as touch-panels have revolutionized the userinterface on mobile phones and GPS navigationdevices, they now are about to change electronicpower in

Page 3 - MODEL 4700 LXI DC LOADS

MODEL 4700 LXI DC LOADSConstant Power operating envelopes100.0 V10.0 V1.0 V0.1 V1 A 10 A 100 A 1,000 A10,000 AAMPSVOLTS3kW6kW18kW24kW36kWCOMPLEX LOAD

Page 4 - SPECIFICATIONS

1Specifications apply at 23o+/- 5oC after a 10 minute warm up.2Accuracies apply when Settings and/or Measurements >10% of Range.3Current linearly r