MODEL 19032/19032-P
Electrical Safety Analyzer
Key Features :
AC/DC/IR/GB/LC five instruments in one
Function test up to 20A
Programmable voltage output and limit
value
OSC open/short check
Flashover detection
Human protection circuit
Multi-scan device support dynamic leakage
current test
Standard RS232 interface
Optional GPIB interface
Large LCD panel
Front panel lockout function
Support PC software
UL/TUV/CE
Key Features 19032-P :
500VA output
Floating output , compliance with EN50191
USB interface, compatible with USB TMC
GFI human body protection circuit
CE certification (only)
General Electrical Safety Testing Solution
Electrical safety testing is one of the major
item in the electrical product quality tests.
All electrical products consisting of adapter,
SMPS, charger, house appliance, information
technology product and video product are
required to perform electrical safety tests.
The 19032 series combines Hi-Pot, IR, GB,
LC/ALC/DLC and Dynamic Function Test. That
can save 50% of production line space without
purchasing several Hi-Pot testers , 19032 is
able to increase efficiency of electrical safety
test during manufacturing and reduce the risk
on testing.
Open/Short Check (OSC)
Patent No. : 254135
All manufacturers have to solve the problems of
error connections and unconnected test cables
caused by the production line testers.Chroma
19032 equipp with the up-to-date open/short
check function (OSC) for product testing.It can
free the tests from such problem.
Twinport
TM
Function
USA Patent No.: US6504381
The key factor affecting the efficiency of
manufacturing is the efficiency for electrical
safety test. Twinport function can lower the time
for safety test, and safety workstation will no
longer be a bottleneck in production line.
Product Application
The 190 32 c an b e a pp lie d t o v er sa til e
tests of electrical products which include
quality assurance sampling inspection test,
manufacturing test and development validation.
• Power cord
Adapter, SMPS
• House appliance
• Information technology product
• Medical equipment
• Lab/testing equipment
• EMI FILTER
EN50191 Floating Output Function
The leakage current of any ground terminal
should be lower than 3.5mA when operating
Floating output function. Therefore, the
operator who near to potential ground terminal
can avoid electrical hazard.
ELECTRICAL SAFETY ANALYZER
MODEL 19032/19032-P
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - MODEL 19032/19032-P

MODEL 19032/19032-PElectrical Safety AnalyzerKey Features : AC/DC/IR/GB/LC five instruments in one Function test up to 20A Programmable voltage

Page 2 - Measurement Technology

OSC function is used to check if the connection is open (bad contact) or short (DUT short) during the test procedure. When circuit opened during the

Page 3

GFI HUMAN PROTECTION (19032-P) FLOATING CIRCUIT FUNCTION (19032-P)For the opoerator to using the Safety Test Instrument more safely, Chroma invente

Page 4 - Product Application

SAFETY TEST OF INFORMATION TECHNOLOGY PRODUCT (IEC 60950) POWER SUPPLIES, ADAPTER AND CHARGER SAFETY TESTINGProduct ApplicationIn digital time, infor

Page 5

SAFETY TEST OF VIDEO PRODUCT (IEC 60065)Video products are inclusive of DVD recorder, LCD TV, audio amplifier, stereo and etc. These products

Page 6 - EST SOFTWARE

Chroma provides related optional test optional xtures for different test requirements . • Isolation Transformer : The isolation transformer is an aux

Page 7 - PANEL DESCRIPTION

Option 19032-P PANEL DESCRIPTION1. Power Switch2. LCD Display3. Stop Key4. Start Key5. Cursor Keys6. TEST Key7. PRINT Key8. SYSTEM Key9. Cal-Enable (1

Page 8 - 致茂電子股份有限公司

19032 : Electrical Safety Analyzer19032-P : Electrical Safety Analyzer (500VA)A190301 : 8HV Scanning BoxA190302 : 5HV/3GC ScannerA190303 : 3HV/5