Vue de la page 0
GE
Sensing
DigitalFlow
DF868
Panametrics Liquid
Ultrasonic Flowmeter
DigitalFlow DF868 is a Panametrics
product. Panametrics has joined
other GE high-technology sensing
businesses under a new name
_
GE Industrial, Sensing.
Applications
The DF868 liquid flowmeter is a complete ultrasonic flow
metering system for measurement of:
Refined hydrocarbons
Petroleum products
Crude oil
Lubricating oils
Diesel and fuel oils
Solvents
Water and wastewater
Hot/chilled water
Water/glycol solutions
Chemicals
Other liquids
Features
Economical, nonintrusive flow measurement
Hazardous (classified) location certification
Simple setup and installation
Suitable for a wide range of pipe sizes and
materials
Two-channel/two-path version available
Energy measurement option
Large backlit LCD display
Velocity, volumetric and energy flow rates
Totalized flow and trend data
g
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - DigitalFlow

GESensingDigitalFlow™DF868Panametrics Liquid Ultrasonic FlowmeterDigitalFlow DF868 is a Panametrics product. Panametrics has joinedother GE high-techn

Page 2

GESensingPanametrics Ultrasonic Liquid FlowmeterThe DigitalFlow DF868 is a full-featured, fixed-installationliquid flowmeter designed to meet all your

Page 3 - Ultrasonic Signal Path

GESensingTransmitsignalTrackingwindowlocked ontoreceive signalTimeVariable timeintervalRange of tracking windowReceivesignalATW ensures accuracy when

Page 4 - Specifications

GESensingDF868SpecificationsOperation and PerformanceFluid TypesAcoustically conductive fluids, including most cleanliquids, and many liquids with ent

Page 5

GESensingDF868SpecificationsStorage Temperature-67°F to 167°F (-55°C to 75°C)Standard Inputs/OutputsTwo 0/4 to 20 mA isolated outputs, 550 S maximum l

Page 6

GESensingDF868SpecificationsTemperature Ranges• Standard: -40°F to 212°F (-40°C to 100°C)• Optional (overall range): -310°F to 1112°F (-190°C to 600°C

Page 7 - Notes In Brief

GESensingDF868ApplicationNotes In BriefClamp-In CapabilityGE offers a wide variety of ways to install transducers toadapt to each piping and applicati

Page 8

GESensing©2005 GE. All rights reserved. 920-001BAll specifications are subject to change for product improvement without notice.DigitalFlow™, Correlat