Fluke i400s Specifications
Specifications
Nominal current
range:
40 A Range 0.5 A to 40 A
400 A Range5 A to 400 A
Basic accuracy:
40 A Range 2% + 0.015 A (45-400 Hz)
400 A Range2% + 0.04 A (45-400 Hz)
Phase Shift:
40 A Range
0.5 to 1 A Unspecified
1 A to 5 A 4°
5 A to 10 A 3°
10 A to 20 A 3°
20 A to 40 A 2°
40 A to 400 A
A
400 A Range
0.5 to 1 A
A
1 A to 5 A
A
5 A to 10 A Unspecified
10 A to 20 A 2°
20 A to 40 A 2°
40 A to 400 A1.5°
Typical Bandwidth:
40 A Range 5 Hz - 10 kHz
400 A Range5 Hz - 10 kHz
Output level(s):
40 A Range 10 mV/A
400 A Range1 mV/A
Crest Factor:
40 A Range ≤ 3
400 A Range≤ 3 to 300 A
≤ 2.5 to 400 A
Safety Specifications
Safety Specifications
Safety CAT IV 600 V, CAT III, 1000 V
Maximum Non-
Destructive Current
1000 V AC
Commentaires sur ces manuels