Product specifications and characteristicsAgilentEPM Series Power MetersE-Series and 8480 Series Power SensorsData Sheet
10The chart in figure 1 shows the typi-cal uncertainty in making a relativepower measurement, using the samepower meter channel and the samepower sens
11E-series E9300 average power sensor specificationsThe E-series E9300 wide dynamicrange, average power sensors aredesigned for use with the EPM fami-
12Typical SWR, 10 MHz to 18 GHz, (25°C ± 10°C) for E9300A and E9301A sensorsTypical SWR, 9kHz to 6 GHz, (25°C ± 10°C) for E9304A sensors
13Typical SWR, 10 MHz to 18 GHz, (25°C ± 10°C) for E9300B and E9301B sensorsTypical SWR, 10 MHz to 18 GHz, (25°C ± 10°C) for E9300H and E9301H sensors
14E-series E9300 average power sensorsTable 10. Power linearity (after zero and cal at ambient environmental conditions)Sensor Power Linearity Lineari
15E-series E9300 average power sensorsTypical E9300B/01B power linearity at 25°C, after zero and calibration, with associatedmeasurement uncertaintyPo
16Note: If the temperature changes after calibration and you choose not to re-calibrate the sensor, the following additional power linearity error sho
17E-series E9300 averagepower sensorsSwitch point dataThe E9300 power sensors have twopaths, a low power path covers: -60 to -10 dBm (“A” suffix senso
18Calibration factor (CF) andreflection coefficient (Rho)Calibration factor and reflectioncoefficient data are provided at frequency intervals on a da
198480 series power sensors(with EPM series powermeters)The 8480 series power sensors aredesigned for use with the 435B,436A, 437B, 438A, 70100A, E141
2EPM series E4418B and E4419B power meters Specifications Specifications describe the instru-ment’s warranted performance andapply after a 30 minute w
208480 series sensors (with EPM series power meters)Table 1525 Watt sensors, 1 mW to 25 W (0 dBm to +44 dBm)Model Frequency range Maximum SWR Power
218480 series sensors (with EPM series power meters)High sensitivity sensors, 100 pW to 10 µW (–70 dBm to –20 dBm)Model Frequency range Maximum SWR
22Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, while minim
3Supplemental characteristicsPower referenceFrequency: 50 MHz nominal.SWR: 1.05 maximum.Connector: Type N (f), 50 Ω.Measurement speedThree measurement
4For E-series sensors:In FAST mode (using free run trigger), within the range –50 dBmto +17 dBm, for a 10 dB decreasingpower step, the settling time i
5For E4412A and E4413A sensorsAuto filter, default resolution, 10 dBdecreasing power step (not across therange switch point)For E-series E9300A sensor
6Power meter functionsAccessed by key entry: either hardkeys,or softkey menu, and programmable.Zero: Zeros the meter. (Power refer-ence calibrator is
7Battery Option 001 operationalcharacteristics1The following information describescharacteristic performance based ata temperature of 25°C unless othe
8EPM series power metersE4418B: Single channel EPM seriespower meterE4419B: Dual channel EPM seriespower meterOptions availablePowerOption E441xB-001:
9E-series power sensorspecificationsThe E-series of power sensors havetheir calibration factors stored inEEPROM and operate over a widedynamic range.